Tel: 0086-25-84041282
Detailed Product Description
dialysis machine :Be applicable of hemodialysis treatment, hemoperfusion, PE, Pure ultrafiltration project..
dialysis machine :Be applicable of hemodialysis treatment, hemoperfusion, Plasma Exchange, Pure ultrafiltration project. Touch acreen was equipped and it can show Na and UF profiles;In addition, the function ie. Backup power supply, Kt/V, HCO3, adjustment, single needle and acetate treatment are optional.
Volume (length width height): 370mm 340mm 1570mm.
Weight: about 90kg
Power supply voltage: AC220V±10% Frequency: 50HZ ~ 60 HZ Power: 1500W
Blood pump/spare pump: Flux: 15 ~ 340ml/min ( 6mm) 20 ~ 460 ml/min ( 8mm)
Heparin pump
Flux: 0.1ml/h10ml/h(±5%)
Precise: 0.1ml/h
Injector size: 20ml/30ml/50ml 10ml/20ml/30ml(selectable)
Arterial pressure
display scope:-300mmHg (±10mmHg)
Venous pressure
display scope:-50mmHg (±10mmHg)
Dialysate flow: 300ml/min800ml/min linearity adjustable (%) (-5%)
Temperature:35.0~39.0
Resolving rate:0.1
Conductivity:13mS/cm~15.5mS/cm
(±0.1 mS/cm)
UF flow scope: 01800ml/h(±30ml/h )
04000ml/h(optional)
ISO UF flow scope:02000ml/h(±30ml/h)
TMP
Scope:-100mmHg
(±20mmHg)
Blood leakage monitor
over 1ml blood per liter dialysate (flow:500ml/min)
Blood level monitor: ultrasonic sensor
Air bubble monitor
Infrared and response threshold value:
Single air bubble of 200 l exists when Blood flux is 200ml/min
Inflow pressure:0MPa0.6Mpa
Inflow temperature: 530
Environment temperature: 1030,
Relative humidity 70%
Rinse/disinfection: chemical disinfection
(citric acid,peracetic acid and oxalic acid)
Hot rinse: 80
Back-up power supply: last for 15-30min after electric-cut..