BF-610TFT Doppler FHR Detector

BF-610TFT Doppler FHR Detector

Features:

  • Portable Fetal Doppler mainly applies to detect the fetal heartbeat rate from the twelfth week;
  • Built-in rechargeable battery can work continuously more than 4 hours;
  • 3.2'' color TFT can clearly display FHR data or graphic , hear rate graphics, battery indicator, volume of sound and synchronous FHR signal;
  • Store and hold FHR data and graphics for up to 12 hours (for option);
  • BF-610P built-in thermal printer can synchronously  print the FHR curve clearly and delicately;
  • Alarm function when FHR is out of the normal range or battery power is in shortage;
  • Probes with high sensitivity can detect FHR and umbilical blood steam and much easier to interchange;
  • Various frequency probes for different use.(for option);
  • Normative collocation: 2.0 MHz
  • RS-232 port can connect the unit with a computer and store the data.
  • Software is option.

 

Specifications:

  • Standard:  EN61266 : 2002
  • Safety Classification: BF/Class Ⅱ/Internal Power.
  • Transducer working frequency: 1.0 MHz (option),2.0 MHz,5.0 MHz(option) ,8.0 MHz (option);(±10%).
  • Ultrasound average intensity: ≤10mW/cm2.
  • Overall Sensitivity: ≥90 dB. The overall Sensitivity of place 200mm away from the surface of probe.  Which is 200mm away from the surface of probe.
  • While detecting the overall Sensitivity, the adopted frequency is (300±50)Hz, the reflecting target speed is 10cm/s~40cm/s.
  • Effective area of transducer: 6.0 ± 0.5 cm2(2MHz).
  • Spatial-peak /temp-peak acoustic pressure:≤ 0.1 MPa.
  • Input Voltage: A.C. 110V~240V  50Hz±1 Hz.
  • Adapter output voltage: DC 24V±2V.
  • Input power:≤ 20VA.
  • Heart rate detecting range: 50bpm —240 bpm.
  • Built-in battery: 14.4V Ni-MH rechargeable battery.
  • Audio maximum output power: ≥1.8 W.
  • Suggested  coupling  impedance: 1.5—1.6(105g/cm2.s).
  • Static continuously working time (when the battery is full of power and it is at the self-detect time): > 8.
  • Doppler Frequency: 100—5000 Hz.
  • Working environment: temperature: +5 ℃ ~ +40 ℃
  • Humidity: ≤ 80%  Atmospheric pressure: 86kPa ~ 106kPa.
  • Transport and storage environment: temperature: -10 ℃ ~ +40 ℃ Humidity: ≤ 80% Atmospheric pressure: 86kPa ~ 106kPa, well-ventilated room without corrosive gases.
  • Acoustic output parameters meet the national exempted conditions which released that the ultrasonic and main unit in all operated mode can meet the following requirements:

    a) The peak negative acoustic pressure (p_) <1Mpa.

    b) The output beam of sound intensity (lob) <20mW/cm2.

    c) Spatial peak time average educed the sound intensity (lspta) <100mW/cm2

Normative collocation: 

BF-610: Main unitAC adapter, 2MHz probe, operation manual, power cable, guarantee card, quality card,  probe boxes.

 

BF-610P: Built-in recorder, two rolls print paper, others are the same with BF-610.