Features:
- Portable Fetal Doppler mainly applies to detect the fetal heartbeat rate from the twelfth week;
- Built-in rechargeable battery can work continuously more than 4 hours;
- 3.2'' color TFT can clearly display FHR data or graphic , hear rate graphics, battery indicator, volume of sound and synchronous FHR signal;
- Store and hold FHR data and graphics for up to 12 hours (for option);
- BF-610P built-in thermal printer can synchronously print the FHR curve clearly and delicately;
- Alarm function when FHR is out of the normal range or battery power is in shortage;
- Probes with high sensitivity can detect FHR and umbilical blood steam and much easier to interchange;
- Various frequency probes for different use.(for option);
- Normative collocation: 2.0 MHz
- RS-232 port can connect the unit with a computer and store the data.
- Software is option.
Specifications:
- Standard: EN61266 : 2002
- Safety Classification: BF/Class Ⅱ/Internal Power.
- Transducer working frequency: 1.0 MHz (option),2.0 MHz,5.0 MHz(option) ,8.0 MHz (option);(±10%).
- Ultrasound average intensity: ≤10mW/cm2.
- Overall Sensitivity: ≥90 dB. The overall Sensitivity of place 200mm away from the surface of probe. Which is 200mm away from the surface of probe.
- While detecting the overall Sensitivity, the adopted frequency is (300±50)Hz, the reflecting target speed is 10cm/s~40cm/s.
- Effective area of transducer: 6.0 ± 0.5 cm2(2MHz).
- Spatial-peak /temp-peak acoustic pressure:≤ 0.1 MPa.
- Input Voltage: A.C. 110V~240V 50Hz±1 Hz.
- Adapter output voltage: DC 24V±2V.
- Input power:≤ 20VA.
- Heart rate detecting range: 50bpm —240 bpm.
- Built-in battery: 14.4V Ni-MH rechargeable battery.
- Audio maximum output power: ≥1.8 W.
- Suggested coupling impedance: 1.5—1.6(105g/cm2.s).
- Static continuously working time (when the battery is full of power and it is at the self-detect time): > 8.
- Doppler Frequency: 100—5000 Hz.
- Working environment: temperature: +5 ℃ ~ +40 ℃
- Humidity: ≤ 80% Atmospheric pressure: 86kPa ~ 106kPa.
- Transport and storage environment: temperature: -10 ℃ ~ +40 ℃ Humidity: ≤ 80% Atmospheric pressure: 86kPa ~ 106kPa, well-ventilated room without corrosive gases.
-
Acoustic output parameters meet the national exempted conditions which released that the ultrasonic and main unit in all operated mode can meet the following requirements:
a) The peak negative acoustic pressure (p_) <1Mpa.
b) The output beam of sound intensity (lob) <20mW/cm2.
c) Spatial peak time average educed the sound intensity (lspta) <100mW/cm2
Normative collocation:
BF-610: Main unitAC adapter, 2MHz probe, operation manual, power cable, guarantee card, quality card, probe boxes.
BF-610P: Built-in recorder, two rolls print paper, others are the same with BF-610.